4.5.1
Personnel Requirements
Personnel Requirements
1
4.5.2
Overview
This procedure can be performed with:
• Fluke tester
4.5.3
Tools and Test Equipment
Item
Fluke ESA612 or equivalent
Note:
The equivalent means that the tool is designed to be capable of performing tests to both the
standard of IEC 62353 and the earth leakage current of IEC 60601-1.
RISK OF ELECTRIC SHOCK.
VOLTAGE PRESENT.
THE SYSTEM MUST BE DISCONNECTED FROM THE FACILITY MAINS INCLUDING THE A1
PANEL GROUND WIRE.
4.5.4
Required Conditions
Condition
All sub-system covers must be removed to access the
grounding points within each sub-system.
4.5.5
Set up tools
• How to connect the Tool
1). Connect power cord
2.) Connect red test lead to red input jack (V/Ω/A)
3.) Connect black test lead to black input jack
6
EVISION
Preliminary Reqs
Not Applicable
Chapter 1 - - Safety and Service
Procedure
30min
Qty.
Effectivity
Part#
1
-
B
XR118 S
RIVO
ERVICE
Finalization
10min
Manufacturer
-
Fluke
Reference
Effectivity
1
M
ANUAL
-
Page 29