Note:
6
EVISION
a.) Load Images
i.) Load the PACS Test Images onto the acquisition workstation. Either use:
DICOM-formatted CDROM with images
Images on system under Image Tools
ii.) Patient Name will appear as GE_PACS_Test_Image, Definium
iii.) Patient ID will be 1000001
b.) Configure PACS to [No-Burn]
i.) Click the Utilities icon
ii.) Click Network Connections – Edit
iii.) Select the Network Host for the PACS to be tested, click Edit
iv.) Click the Preferences tab
v.) Make sure that Apply Burn-on-Send to images... is not checked
c.)
Send No-Burn image
i.) Highlight PACS Test Image #1 in the Images list
All images in the GE_DX_Image_Display_Test exam are named "1". In the
following steps, when an Image number (1, 2, 3....) is called out, you should
select that image ' position' from the Images list. For example: image number 1
is the first image on the list, image 6 is the sixth image on the list. To verify that
you have selected the correct image, the image number can be viewed in the
upper left corner of the image.
ii.) Click the appropriate PACS Network Host in the Destination icon list
d.) Configure PACS to Burn-on-Send
i.) Click the Utilities icon, click Network Connections – Edit
ii.) Select the Network Host for the PACS to be tested, click Edit
iii.) Click the Preferences tab; check the option Apply Burn-on-Send to images...
e.) Send Burn-on-Send image
Chapter 6 - - Calibration
B
XR118 S
M
RIVO
ERVICE
ANUAL
Page 149