GE Healthcare Brivo XR118 Service Manual page 290

Table of Contents

Advertisement

GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Parameter
Grid Status
TE Settings
12. When finished adjusting parameters, click [Save] or [Save To Multiple].
 [Save] applies the parameters to the currently selected patient size.
 [ Save To Multiple ] applies the parameters to multiple patient sizes within the
13. Click [Close] on the Image Processing Preference Editor.
14. Log off and reset the system for changes to take effect. Refer to
Installation & Configuration
Page 290
6
EVISION
If grid status is <Out>, Grid Line Removal algorithm will be not
used. If grid status is <In> and at the same time Grid Line
Reduction selection in Utility is On, Grid line reduction algorithm
will be used.
See Tissue Equalization section for range and effect of TE sliders.
Refer to
Tissue Equalization Overview
Note:
The combined total of the TE Under-penetrated and TE Over-penetrated
areas cannot exceed 100%. The slider will automatically stop and a
message appears: "The Total Area cannot exceed 100%. Reduce the
(Under- or Over-) Penetrated area to proceed."
– A message appears: "Save changes?"
– Click [Yes].
selected anatomy.
– The Save to Multiple screen appears.
Patient Size
– Select the
– Click [Save].
Chapter 12 - - Set Preferences
Definition
for more information.
to save the new look to.
for more information.
B
XR118 S
M
RIVO
ERVICE
ANUAL