Chapter 6 - Calibration
Section 1.0

Before You Begin

1.1
Service User Interface (SUIF)
A list of available system calibrations is displayed on the Service User Interface, whenever the
calibration tab is chosen.
Be patient when executing commands. Text status messages are generally displayed in status line,
whenever a calibration icon is selected. If you are uncertain whether a command is being executed,
check the status line first. Repeated attempts to execute a command may require you to exit the
SUIF and re-enter to gain control again.
In calibration and QAP page of SUIF, if you want to exit the page when an operation is in progress,
the "cancel" button should be clicked before clicking the exit button.
1.2
Calibration Sequence
Calibrations must be performed in the order shown on the left hand side of the calibration tab. The
calibration summary screen only displays the date that a specific calibration was last performed.
Calibrations listed on the summary are not listed in the order they should be performed.
1.3
Returning the System to the User
1.) After completing calibrations, delete any New Patients files created during service, and before
2.) Perform a system reset. A system reset must be performed to invoke changes. Application
3.) Perform a system backup following calibration and after the system has reset. In order to
6
EVISION
Figure 6-1 Service User Interface (SUIF) ICONs
turning the system over to the end-user.
software and most SUIF tools only reads the system calibration files during boot-up. If you
make changes, they will not be invoked until you reset (reboot) the system (See Invoking
Calibration Changes).
preserve any new calibrations, a system backup must be made.
Chapter 6 - - Calibration