GE H
EALTHCARE
D
5461425-8EN, R
IRECTION
Section 2.0
Calibration Sequence
Potential for Data Loss
1.) Always create an up-to-date backup of your system data following product installation or after
2.) The system must be reset, upon exiting the Service User Interface for Configuration and
Note:
Table 6-1 First Installation Calibration
Tool
Detector Calibration
Perform System Reset
Image Monitor Setup
PACS / Image Display Test Image
Quality Assurance Procedure (QAP)
* Numbers in column indicates the order/sequence that the task is performed.
You must always calibrate your system before it can be used and remember to restart your system
after the calibration to active the calibration.
Do backup after calibrations in System Data Backup and Restore
Section 3.0
Invoking New Calibrations
Please restart the system to invoke calibration changes.
The system must be rebooted if changes are made to calibrations files. This includes the creation
of new calibrations and/or the restoration of calibration files. Applications software running on this
system only reads calibration files during workstation boot-up. Any changes after boot-up are not
detected by the applications software.
Section 4.0
Saving Changes
Save changes to calibration files by performing a backup. See System Data Backup and Restore
6
EVISION
making any calibrations.
Calibration information to take affect. This must be done before the system is used with patient
acquisitions.
After first installation, those calibrations are needed. Follow the steps below.
Chapter 6 - - Calibration
Calibration/Task
Detector
Image Monitor Setup with Flat Panel
Monitor
PACS / Image Display Test Image
QAP
B
XR118 S
M
RIVO
ERVICE
ANUAL
Order
1
2
3
4
5
Page 137