Ion Chamber Detectors; Applications For Detector Sensing Areas; Areas 1 And 3 - GE Definium 5000 Operator's Manual

X-ray system
Table of Contents

Advertisement

Ion Chamber Detectors

Ion Chamber Detectors have three sensing areas (Figure 7-3). Sensing Area 2 is located at the center of
the X-ray beam. While Area 1 and Area 3 can be selected to cover an exposure of two symmetrical parts
of the body, such as the lungs or the kidneys, care should be taken to center the patient and detector
areas accordingly.
Figure 7-3 Sensing areas
The position of the sensing areas are shown in relation to the area of a 210 mm x 248 mm (8.25in x
9.75in) Collimator Light Field.

Applications for Detector Sensing Areas

The detector sensing areas should be used as described in the following sections. You should become
familiar with their locations and recommended use. The sensing areas are numbered 1 through 3.

Areas 1 and 3

Areas 1 and 3 are used to cover symmetrical body parts. For example, acquiring a chest radiograph
includes the lungs, which are proportional parts of the body. In this application, Area 1 and 3 must be
located in line with radiation transmitted through the left and right lung fields. This ensures these areas
are not influenced by variations in tissue opacity caused by the heart or vertebrae.
Definium 5000 X-Ray System
Operator Manual
Chapter 7: Automatic Exposure Control (AEC)
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
7-4