Maximum Allowance Limit For Probe Leakage Current; Data Sheet For Transducer Source Leakage Current - GE RTfino/RT3200 Service Manual

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 21
3–7–4 Probe Leakage Current Test (Continued)

Data Sheet for Transducer Source Leakage Current

Follow the foregoing test procedure for every transducer. The test passes when all readings measure less than the
value shown in the table below (TABLE 3–5).
1. Type BF Applied Part Leakage Current Limits
[Non–conductive (floating) Surface and Cavity Transducers]
COUNTRY
USA
OTHERS
2. Type CF Applied Part Leakage Current Limits
[Surgical Transducer and ECG Connections]
COUNTRY
USA
OTHERS
Transducer tested:
RTfino/RT3200 Advan-
tage-II/RT-x200 Unit Power
ON
ON
ON
ON
OFF
OFF
OFF
OFF
RTfino/RT3200 Advantage–II/RT–x200 SERVICE MANUAL
TABLE 3–5

MAXIMUM ALLOWANCE LIMIT FOR PROBE LEAKAGE CURRENT

NORMAL
OPEN
CONDITION
GROUND
0.05mA
0.05mA
0.1mA
0.5mA
NORMAL
OPEN
CONDITION
GROUND
0.01mA
0.01mA
0.01mA
0.05mA
TABLE 3–6
DATA SHEET FOR TRANSDUCER SOURCE LEAKAGE CURRENT
Tester Power
Polarity
NORM
NORM
REV
REV
NORM
NORM
REV
REV
3–16
REVERSE
OPEN
POLARITY
NEUTRAL
0.05mA
0.05mA
0.5mA
0.5mA
REVERSE
OPEN
POLARITY
NEUTRAL
0.01mA
0.01mA
0.05mA
0.05mA
Tester Ground
Switch
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
OPEN
CLOSED
INSTALLATION AND REPLACEMENT
P9030GX
Measurement