:SBUS<n> Commands
:SBUS<n>:A429:AUTosetup
Command Syntax
:SBUS<n>:A429:AUTosetup
The :SBUS<n>:A429:AUTosetup command automatically sets these options
for decoding and triggering on ARINC 429 signals:
• High Trigger Threshold: 3.0 V.
• Low Trigger Threshold: - 3.0 V.
• Noise Reject: Off.
• Probe Attenuation: 10.0.
• Vertical Scale: 4 V/div.
• Serial Decode: On.
• Base (:SBUS<n>:A429:BASE): HEX.
• Word Format (:SBUS<n>:A429:FORMat): LDSDi (Label/SDI/Data/SSM).
• Trigger: the specified serial bus (n of SBUS<n>).
• Trigger Mode (:SBUS<n>:A429:TRIGger:TYPE): WSTArt.
Errors
See Also
636
(see
page
1138)
"- 241, Hardware missing"
":SBUS<n>:A429:BASE"
":SBUS<n>:A429:FORMat"
":SBUS<n>:A429:TRIGger:TYPE"
"Introduction to :SBUS<n> Commands"
":SBUS<n>:MODE"
on page 633
":SBUS<n>:A429 Commands"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 1099
on page 637
on page 641
on page 650
on page 629
on page 634