Measure:ipower - Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

:MEASure:IPOWer

Command Syntax
:MEASure:IPOWer
The :MEASure:IPOWer command installs an input power measurement on
screen.
Before sending this command or query, you must specify the channels
probing the input voltage, input current, output voltage, and output
current (using the :POWer:SIGNals:SOURce:VOLTage<i> and
:POWer:SIGNals:SOURce:CURRent<i> commands) and you must perform the
automated signals setup (using the :POWer:SIGNals:AUTosetup EFFiciency
command).
Query Syntax
:MEASure:IPOWer?
The :MEASure:IPOWer query returns the measured input power.
Return Format
<return_value><NL>
<return_value> ::= the input power value in NR3 format
See Also
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
":POWer:SIGNals:SOURce:VOLTage<i>"
":POWer:SIGNals:SOURce:CURRent<i>"
":POWer:SIGNals:AUTosetup"
":POWer:EFFiciency:APPLy"
:MEASure Power Commands
on page 582
on page 581
on page 564
on page 535
21