:POWer:SIGNals:IEXPected
Command Syntax
:POWer:SIGNals:IEXPected <value>[suffix]
<value> ::= Expected current value in NR3 format
[suffix] ::= {A | mA}
The :POWer:SIGNals:IEXPected command specifies the expected inrush
current amplitude. This value is used to set the vertical scale of the
channel probing current.
Query Syntax
:POWer:SIGNals:IEXPected?
The :POWer:SIGNals:IEXPected query returns the expected inrush current
setting.
Return Format
<value><NL>
<value> ::= Expected current value in NR3 format
See Also
•
•
•
•
•
•
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
(see
page
1138)
":POWer:INRush:APPLy"
":POWer:SIGNals:AUTosetup"
":POWer:SIGNals:OVERshoot"
":POWer:SIGNals:VMAXimum:INRush"
":POWer:SIGNals:SOURce:CURRent<i>"
":POWer:SIGNals:SOURce:VOLTage<i>"
on page 547
on page 564
on page 574
on page 575
on page 581
on page 582