20
:MEASure Commands
:MEASure:NPULses
Command Syntax
:MEASure:NPULses [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:NPULses command installs a falling pulse count
measurement on screen. If the optional source parameter is not specified,
the current source is measured.
This command is not available if the source is FFT (Fast Fourier Transform).
N O T E
Query Syntax
:MEASure:NPULses? [<source>]
The :MEASure:NPULses? query measures and returns the on- screen falling
pulse count.
Return Format
<value><NL>
<value> ::= the falling pulse count in NR3 format