27
:SBUS<n> Commands
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA
Command Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA <string>
<string> ::= "nn...n" where n ::= {0 | 1 | X | $}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA command defines the SPI
data pattern resource according to the string parameter. This pattern,
along with the data width, control the data pattern searched for in the
data stream.
If the string parameter starts with "0x", it is a hexadecimal string made up
of hexadecimal and X (don't care) characters; otherwise, it is a binary
string made up of 0, 1, and X (don't care) characters.
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh should be set before
N O T E
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA.
Query Syntax
:SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA?
The :SBUS<n>:SPI:TRIGger:PATTern:MOSI:DATA? query returns the current
settings of the specified SPI data pattern resource in the binary string
format.
Return Format
<string><NL>
See Also
•
•
•
750
(see
page
1138)
"Introduction to :TRIGger Commands"
":SBUS<n>:SPI:TRIGger:PATTern:MOSI:WIDTh"
":SBUS<n>:SPI:SOURce:MOSI"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 877
on page 751
on page 747