33
:WGEN Commands

:WGEN:ARBitrary:INTerpolate

Command Syntax
:WGEN:ARBitrary:INTerpolate {{0 | OFF} | {1 | ON}}
The :WGEN:ARBitrary:INTerpolate command enables or disables the
Interpolation control.
Interpolation specifies how lines are drawn between arbitrary waveform
points:
• When ON, lines are drawn between points in the arbitrary waveform.
• When OFF, all line segments in the arbitrary waveform are horizontal.
Query Syntax
:WGEN:ARBitrary:INTerpolate?
The :WGEN:ARBitrary:INTerpolate query returns the current interpolation
setting.
Return Format
{0 | 1}
See Also
1004
(see
page
1138)
Voltage levels change linearly between one point and the next.
The voltage level of one point remains until the next point.
":WGEN:ARBitrary:DATA"
":WGEN:ARBitrary:DATA:DAC"
":SAVE:ARBitrary[:STARt]"
":RECall:ARBitrary[:STARt]"