:MEASure Power Commands

:MEASure:OFFTime

Command Syntax
:MEASure:OFFTime [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:OFFTime command installs an "off time" measurement on
screen.
Turn off time measures the difference of time between when the input AC
Voltage last falls to 10% of its maximum amplitude to the time when the
output DC Voltage last falls to 10% of its maximum amplitude.
The <source1> parameter is the AC Voltage and the <source2> parameter is
the DC Voltage. These sources can also be specified by the
:MEASure:SOURce command.
Query Syntax
:MEASure:OFFTime? [<source1>][,<source2>]
The :MEASure:OFFTime query returns the measured turn off time.
Return Format
<return_value><NL>
<return_value> ::= the time in seconds in NR3 format
See Also
480
(see
page
1138)
":MEASure:SOURce"
":POWer:ONOFf:TEST"
":POWer:ONOFf:APPLy"
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 444
on page 556
on page 553