:MTESt Commands

:MTESt:RMODe

Command Syntax
:MTESt:RMODe <rmode>
<rmode> ::= {FORever | SIGMa | TIME | WAVeforms}
The :MTESt:RMODe command specifies the termination conditions for the
mask test:
• FORever — the mask test runs until it is turned off.
• SIGMa — the mask test runs until the Sigma level is reached. This level
• TIME — the mask test runs for a fixed amount of time. The amount of
• WAVeforms — the mask test runs until a fixed number of waveforms are
Query Syntax
:MTESt:RMODe?
The :MTESt:RMODe? query returns the currently set termination condition.
Return Format
<rmode><NL>
<rmode> ::= {FOR | SIGM | TIME | WAV}
See Also
Example Code
508
(see
page
1138)
is set by the
":MTESt:RMODe:SIGMa"
time is set by the
":MTESt:RMODe:TIME"
acquired. The number of waveforms is set by the
":MTESt:RMODe:WAVeforms"
"Introduction to :MTESt Commands"
":MTESt:RMODe:SIGMa"
":MTESt:RMODe:TIME"
":MTESt:RMODe:WAVeforms"
"Example Code"
on page 491
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 513 command.
on page 514 command.
on page 515 command.
on page 491
on page 513
on page 514
on page 515