Waveform Invert

Press 1 to open the setting menu of CH1. Then, press Invert continuously to enable
or disable the waveform invert. When enabled, the channel label is displayed, as
shown in the following figure.
When disabled, the waveform is displayed normally; when enabled, the voltage
values of the displayed waveform are inverted (as shown in Figure 2-1). Enabling the
waveform invert will also change the result of math function and waveform
measurement.
(a) "Invert" Off
Note:
When the waveform invert is enabled, the trigger edge or the trigger polarity
will change (e.g. Edge trigger, Pulse trigger, or Slope trigger).

To Set the Probe

The analog channel of this oscilloscope not only supports the common passive probe,
but also the active probe. It can automatically recognize the currently connected
probe type and its probe ratio. For detailed technical information of the probes,
please refer to the corresponding Probe User's Guide.
Press 1 to open the setting menu of CH1. Then, press Probe to open the probe
menu. If different probes are connected to the oscilloscope, different information
2-8
Figure 2-1 Waveform Invert
Chapter 2 To Set the Vertical System
(b) "Invert" On
MSO7000/DS7000 User Guide