Slit Width - Agilent Technologies G1315B User Manual

Agilent 1200 series diode array and multiple wavelength detectors
Table of Contents

Advertisement

Figure 30

Slit Width

The detector has a variable slit at the entrance of the spectrograph. This is an
effective tool to adapt the detector to changing demand of different analytical
problems.
A narrow slit provides spectral resolution for analytes with very fine
structures in the absorbance spectrum. An example of such a spectrum is
benzene. The five main absorbance bands (fingers) are only 2.5 nm wide and
just 6 nm apart from each other.
1200 Series DAD and MWD User Manual
Grad.: 0.02 m KH
PO
/ACN from 12% ACN to 45% ACN in 12 min
2
4
Gradient Analysis of PTH-Amino Acids (1 pmol each), with and without Refer-
ence
Time (min)
How to optimize the Detector
1 pmol each
Wavelength
267 nm
Reference 380 nm
Wavelength
267 nm
No reference
5
77