Appendix | ASCII Communication with Controller
A 5.3.15.4 Set Scaling for DAC
ANALOGSCALEMODE STANDARD | TWOPOINT
Selects whether to use one-point or two-point scaling of the analog output.
- STANDARD --> One-point scaling
- TWOPOINT --> Two-point scaling
The standard scaling is configured for distances -MB/2 to MB/2 and for thickness measurement from 0 to 2 MB
(MB=measuring range).
Minimum and maximum measured values must be specified in millimeters. The available output range of the analog
output is then spread between the minimum and maximum measured values. The minimum and maximum measured
values must be between -2147.0 and 2147.0.
The minimum and maximum measured values are processed with three decimal places.
Command is mapped in the SDO 0x31D0.
A 5.3.15.5 Set Scaling Range
ANALOGSCALERANGE <limit 1> <limit 2>
Two-point scaling requires the start and end of the range to be entered in millimeters.
- <limit 1> = (-2147.0 ... 2147.0) [mm], and different from <limit 2>.
- <limit 2> = (-2147.0 ... 2147.0) [mm], and different from <limit 1>.
The values cannot be identical.
Command is mapped in the SDO 0x31D0.
A 5.3.16 System Settings
A 5.3.16.1 Key Lock
KEYLOCK NONE | ACTIVE | (AUTO [<value>])
Selection of the key lock.
- NONE: Key always functions; no key lock
- ACTIVE: Key lock activates immediately upon restart
- AUTO: Key lock is only activated <time> minutes after restart, value range 1 ... 60 min
Command is mapped in the SDO 0x34A0.
A 5.3.16.2 Switch EtherCAT to Ethernet Setup Mode
BOOTMODE [FIELDBUS|RECOVERY]
Switches firmware. No action is taken if the requested firmware is already active. Otherwise, the requested firmware is
installed and a restart takes place. The switch takes approx. one minute. The power supply for the IFD must not be inter-
rupted during this time.
- FIELDBUS: Start with EtherCAT
- RECOVERY: Start in Ethernet setup mode
confocalDT 2410/2411/2415