7.8.3 External adjustment with user-defined adjustment weight < caleud >
22
 Place the adjustment weight.
 < wait> followed by < reMvld> will be displayed.
 Once < reMvld> is displayed, remove the adjustment
weight.
 After successful adjustment the balance automatically re-
turns to weighing mode.
In case of an adjustment error (e.g. objects on the weigh-
ing plate) the display will show the error message <
wrong >. Switch off balance and repeat the adjustment
process.
 Press and hold the TARE and ON/OFF buttons simulta-
neously to enter the setup menu.
 Wait until the first menu item < Cal > is displayed.
 Confirm by [  ] button, < CalExt> will be displayed.
 Use the navigation keys to select   < caleud>.
TEWJ-B-BA-e-2310