13 Description of individual menu items

13.1 Automatic cut-off of display backlight

(Auto backlight OFF)
Menu item: [
] see chpt. 10.1.2
B. AB
To save battery when function is enabled backlight of display will automatically
switch off after 3 minutes during which there was no change of load or activity.

13.2 How to change weighing units

Menu item: [
Cq W.A
This setting can be used to set different display units (A or B) for a weighing value.
Use the F-key to choose between units A and B.
• "Units A" are available in all operating modes.
• "Units B" are only available in weighing mode [

13.3 How to change readability (models AES only)

Menu item: [
This function is used to change readability for weighing unit A or B. The last decimal
place will be rounded or removed from the display.
Example for weighing unit [ g ] :
Function 1
Menu settings
AES/AEJ-BA-e-0912
] and [
] see chpt. 10.1.2.
Cs W.B
] and [
] see chpt. 10.1.2.
Readability
1
0.0001 g
2
0.0002 g
3
0.0005 g
4
0.001 g
5
0.002 g
q SET 1
].
Reaction to ambient con-
ditions
sensitive
insensitive