11.4 Changing readability

(only models ABT 100-5M, ABT 120-5DM, ABT 220-5DM)
Models ABT 120-5DM and ABT 220-5DM are set to "lower range" with a readability
of 0.01 mg as soon as they are connected to power supply and turned on. To switch
to the "upper range" with a minimum display of 0.1mg, press the [1d/10d] key.
If the lower max. weighing range (82g for ABT 220-5DM, 42 g for ABT 120-5DM) is
exceeded during the weighing process, the balance changes automatically to the
upper range.
If you tare the balance in the upper range, this setting will be maintained throughout.
To return to lower range, you will have to actuate the [TARE] key after the stability
symbol ( ) has appeared.
36
Lower range:
Upper range:
ABT-BA-e-1312