Piece Counting; Counting With Reference Quantity 5, 10 Or 20 - KERN CKE 6K0.02 Operating Instructions Manual

Hide thumbs Also See for CKE 6K0.02:
Table of Contents

Advertisement

Available languages

Available languages

9.2 Piece counting

Before the balance can count parts, it must know the average piece weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
• The higher the reference quantity the higher the counting exactness.
• Especially high reference must be selected for small parts or parts with
considerably different sizes.
• Smallest counting weight see table „Technical data".

9.2.1 Counting with reference quantity 5, 10 or 20

The self-explanatory control panel visualises the sequence of the required steps:
Put the empty container on the weighing plate and press the TARE
button. The container is tared, the zero display will appear.
Fill with the container reference parts (e.g. 5, 10 or 20 pieces).
Confirm the reference quantity selected by pressing the key (5x, 10x,
20x). The balance will calculate the average item weight and then
displays the quantity of parts.
Remove reference weight. The balance is now in piece counting mode
counting all units on the weighing plate.
Fill the counting quantity. The piece quantity is shown directly in the
display.