Program Overall - Agilent Technologies ENA Series Service Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Step 10. To exit the program, click "Exit" button.

Program Overall

Instrument
Information
Test Conditions
Test Items
Figure 2-6 Test dialog box
Chapter 2
Instruction for Using the Performance Test Software
The option and serial number for the E5061A/E5062A, and test date
can be entered automatically.
The test conditions, "Temperature", "Humidity" and person doing the
test ("Tested by") can be entered.
The following test can be selected. To execute the test you select, click
"Execute" button.
Frequency Accuracy Test
RF Output Level Accuracy Test
RF Output Level Linearity Test
Trace Noise Test
Crosstalk & System Dynamic Range Test
Dynamic Accuracy (Magnitude) Test
Uncorrected System Performance Test
When each test is executed, the test dialog box is displayed. The
following screen is an example of the test dialog box.
To execute the test, click "Execute" button. To return the
E5061A/E5062A performance test program dialog box (Main Menu),
click "Back to Main Menu".
The test program will prompt you to connect cables and O/S/L
terminations to the equipment or E5061A/E5062A. Follow the
instructions as shown below on the E5061A/E5062A display for
performing the test.
Performance Tests
31