Uncorrected System Performance Test - Agilent Technologies ENA Series Service Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Performance Tests
E5062A Performance Test Record (for Option 150/250)
Power level [dBm]
Test limit [dB]
0
-5
-15
-20
-25
-30
-35
-40
-45
-50
-55
-60
-65
-70
-75
-80
-85
-90
-95
-100

Uncorrected System Performance Test

@ Correction: Off
S21 or S11
System
Frequency range [Hz]
performance
Directivity
Directivity
Directivity
Source match
66
Test result [dB]
 0.090
 0.068
 0.065
 0.073
 0.080
 0.087
 0.095
 0.103
 0.111
 0.121
 0.134
 0.150
 0.174
 0.211
 0.270
 0.371
 0.546
 0.841
 1.334
2.141
Test limit [dB]
300 k - 100 M
100 M - 1.5 G
1.5 G - 3 G
300 k - 100 M
Measurement
uncertainty [dB]
 0.011
 0.011
 0.011
 0.011
 0.015
 0.015
 0.019
 0.019
 0.022
 0.022
 0.024
 0.024
 0.026
 0.026
 0.028
 0.028
 0.030
 0.030
 0.032
 0.032
Test result [dB]
 -25
 -25
 -25
 -25
Chapter 2