Performance Tests
E5062A Performance Test Record (for Option 150/250)
Power level [dBm]
Test limit [dB]
0
-5
-15
-20
-25
-30
-35
-40
-45
-50
-55
-60
-65
-70
-75
-80
-85
-90
-95
-100
Uncorrected System Performance Test
@ Correction: Off
S21 or S11
System
Frequency range [Hz]
performance
Directivity
Directivity
Directivity
Source match
66
Test result [dB]
0.090
0.068
0.065
0.073
0.080
0.087
0.095
0.103
0.111
0.121
0.134
0.150
0.174
0.211
0.270
0.371
0.546
0.841
1.334
2.141
Test limit [dB]
300 k - 100 M
100 M - 1.5 G
1.5 G - 3 G
300 k - 100 M
Measurement
uncertainty [dB]
0.011
0.011
0.011
0.011
0.015
0.015
0.019
0.019
0.022
0.022
0.024
0.024
0.026
0.026
0.028
0.028
0.030
0.030
0.032
0.032
Test result [dB]
-25
-25
-25
-25
Chapter 2