Agilent Technologies ENA Series Service Manual page 67

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

E5062A Performance Test Record (for Option 150/250)
System
Frequency range [Hz]
performance
Source match
Source match
Load match
Load match
Load match
Reflection Tracking
Reflection Tracking
Reflection Tracking
Transmission
Tracking
Transmission
Tracking
Transmission
Tracking
S12 or S22 (Option 250 only)
System
Frequency range [Hz]
performance
Directivity
Directivity
Directivity
Source match
Source match
Source match
Load match
Load match
Load match
Reflection Tracking
Reflection Tracking
Reflection Tracking
Transmission
Tracking
Chapter 2
Test limit [dB]
 -25
100 M - 1.5 G
 -25
1.5 G - 3 G
 -15
300 k - 100 M
 -15
100 M - 1.5 G
 -15
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
 1.0
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
 1.0
1.5 G - 3 G
Test limit [dB]
 -25
300 k - 100 M
 -25
100 M - 1.5 G
 -25
1.5 G - 3 G
 -25
300 k - 100 M
 -25
100 M - 1.5 G
 -25
1.5 G - 3 G
 -15
300 k - 100 M
 -15
100 M - 1.5 G
 -15
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
 1.0
1.5 G - 3 G
 1.0
300 k - 100 M
Performance Tests
Test result [dB]
Test result [dB]
67