Front Panel Removal

Tools required

TORX screwdriver T15 and T20
Flat edge screwdriver

Procedure

Refer to Figure 6-2 for this procedure.
Step 1. Remove the outer cover as described in "Outer Cover Removal" on page 166.
Step 2. Release the cables from the cable clamp (item 1).
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the eight TORX T15 screws (item 3) fastening front panel.
Step 5. Gradually push the front panel assembly towards the outside.
Step 6. Release the mylar cable and flat cables from the cable clamps (item 6).
Step 7. Disconnect the three cables (item 4, 5 and 8) from the front panel.
NOTE
Keep the mylar cable (item 7) connected.
The mylar cable and its connector are very fragile. It's recommended to replace the mylar
cable when the cable is disconnected and connected frequently.