Agilent Technologies ENA Series Service Manual page 79

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

E5062A Performance Test Record (for Option 175/275)
System
Frequency range [Hz]
performance
Load match
Load match
Load match
Reflection Tracking
Reflection Tracking
Reflection Tracking
Transmission
Tracking
Transmission
Tracking
Transmission
Tracking
S12 or S22 (Option 275 only)
System
Frequency range [Hz]
performance
Directivity
Directivity
Directivity
Source match
Source match
Source match
Load match
Load match
Load match
Reflection Tracking
Reflection Tracking
Reflection Tracking
Transmission
Tracking
Transmission
Tracking
Chapter 2
Test limit [dB]
 -15
300 k - 100 M
 -15
100 M - 1.5 G
 -15
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
 1.0
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
 1.0
1.5 G - 3 G
Test limit [dB]
 -25
300 k - 100 M
 -25
100 M - 1.5 G
 -25
1.5 G - 3 G
 -25
300 k - 100 M
 -25
100 M - 1.5 G
 -25
1.5 G - 3 G
 -15
300 k - 100 M
 -15
100 M - 1.5 G
 -15
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
 1.0
1.5 G - 3 G
 1.0
300 k - 100 M
 1.0
100 M - 1.5 G
Performance Tests
Test result [dB]
Test result [dB]