Table 7-1
Post-Repair Procedures
Replaced
Required Adjustments
Assembly or Part
Correction Constants (CC)
A24 GPIB Card
No adjustment needed
A27 HDD
Perform "HDD" (Hard Disk Drive) in
Spot Adjustment of the program.
Calibration of the Touch Screen (for opt.
016)
A28 FDD
No adjustment needed
A50 Power Supply
No adjustment needed
Assembly
A51 LCD
Not adjustment needed
Touch panel (for
Calibration of the Touch Screen
Opt. 016)
A52 Inverter Board
No adjustment needed
Chapter 7
Post-Repair Procedures
Post-Repair Procedures
Verification
"To Check the GPIB" on page 116
Inspect the Booting Process
"To Check the FDD" on page 111
Inspect the Booting Process
Inspect the Booting Process
"To Check the LCD" on page 110
Inspect the Booting Process
"To Check the Touch Panel (Option 016
only)" on page 110
Inspect the Booting Process