8. This screen allows you to set the Type of cabinet.
Leave this setting alone, and press
9. This screen allows you to configure the cabinet for an
airflow sensor or not. If you are not changing the
selection, press [OK/MUTE].
10. This screen configures the unit for a UV light or not.
Leave this setting alone, and press
11. This screen allows you to configure the cabinet for a
canopy or not. Use
After your selection is chosen, press [OK/MUTE].
12. This screen allows you to configure the cabinet's sash
height. Leave this setting alone, and press [OK/MUTE].
The unit will then return to the Home Screen in the new
configuration.
13. Take Inflow and Downflow readings as described in
within acceptable tolerance for the model, adjust the biosafety cabinet's internal
damper and/or blower(s) speed until Inflow and Downflow are correct. Use the
original blower speed(s) as a reference.
[OK/MUTE].
[OK/MUTE].
[UP]
or
[DOWN]
to select Canopy.
170
TYPE C1
Section 9:
Calibration. If not