Exhaust System Requirements
The exhaust system must be capable of moving the following volumes of exhaust air at
the negative pressures listed. The Airflow Volumes are the values recorded via direct
measurement using a flow hood at the front opening of the cabinet. The Concurrent
Balance Values are measured in the exhaust duct via traverse methodology, and will
always be higher due to differences in volume measurement methodologies.
For Type A2 models with Ventus Canopy Connections, see Table 7-1.
For Cell Logic+ Type A2 models with Ventus Canopy Connections, see Table 7-2
For Type B2 models, see Table 7-3.
For Type C1 models operating in B-mode, see Table 7-4.
Type A2 Models
Logic+ Type A2 Model
3-foot, 8" Sash
3-foot, 10" Sash
4-foot, 8" Sash
4-foot, 10" Sash
4-foot, 12" Sash
5-foot, 8" Sash
5-foot, 10" Sash
5-foot, 12" Sash
6-foot, 8" Sash
6-foot, 10" Sash
6-foot, 12" Sash
1: Unlike Type B biosafety cabinets, the recommended vacuum will remain constant
throughout the life of the exhaust HEPA filter. Duct vacuums below 0.05 inches H
(12 Pa) or above 0.5 inches H
alarm condition from the product.
2: WC = Inches of Water Column, typically expressed in units of inches H
Airflow Volume
3
3
ft
/min
m
/hr
ft
263
447
316
537
333
566
404
687
474
805
403
685
491
834
580
985
473
804
579
984
686
1166
O (125 Pa) may result in erratic operation and throw an
2
31
Concurrent
Recommended
Balance Value
Duct Vacuum
3
3
/min
m
/hr
289
491
348
591
366
622
444
754
521
885
443
753
540
917
638
1084
520
883
637
1082
755
1283
Table 7-1
1
2
WC
Pa
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
0.15
38
O
2
O.
2
Need help?