Cell Logic+ Type A2 Models; Type B2 Models - Labconco Logic+ A2 Technical Manual

Biosafety cabinets, procedure stations
Hide thumbs Also See for Logic+ A2:
Table of Contents

Advertisement

Cell Logic+ Type A2 Models

Logic+ Type A2 Model
3-foot, 9" Sash
4-foot, 9" Sash
5-foot, 9" Sash
6-foot, 9" Sash
1: Unlike Type B biosafety cabinets, the recommended vacuum will remain constant
throughout the life of the exhaust HEPA filter. Duct vacuums below 0.05 inches H
(12 Pa) or above 0.5 inches H
alarm condition from the product.
2: WC = Inches of Water Column, typically expressed in units of inches H

Type B2 Models

Logic+ Type B2 Model
4-foot, 8" Sash
6-foot, 8" Sash
1: WC = Inches of Water Column, typically expressed in units of inches H
Airflow Volume
3
3
ft
/min
m
/hr
ft
290
492
369
627
447
760
526
894
O (125 Pa) may result in erratic operation and throw an
2
Airflow Volume
3
ft
/min
m
3
/hr
ft
723
1228
1083
1840
32
Concurrent
Recommended
Balance Value
Duct Vacuum
3
3
/min
m
/hr
319
541
405
688
492
835
579
983
Concurrent
Recommended
Balance Value
Duct Vacuum
3
/min
m
3
/hr
852
1448
1265
2149
Table 7-2
1
2
WC
Pa
0.15
38
0.15
38
0.15
38
0.15
38
O
2
O.
2
Table 7-3
1
WC
Pa
1.8
450
2.2
550
O.
2