15: Diagnostics

This section details diagnostic functions available through the software of the biosafety
cabinet.

Sash Sensors

1. From the Home Screen, press [MENU].
2. Press
[UP]
to select the
3. Press
[OK/MUTE]
4. Press
[DOWN]
highlighted, then
5. The
Sash Sensors
[OK/MUTE]
Note: If the blower is on when entering the Sash
Sensor Diagnostic screen, it will automatically
turn off.
6. Slowly raise the sash from fully closed to open. When
closed, the screen will report CLOSED. Below or
above the proper sash height, the screen will report
NOT AT WORKING
height (+/- 0.75 inches), the screen will report
WORKING
HEIGHT. Type A2 12-inch and C1 will also
report
ALMOST CLOSED
cm) above closed.
7. Press
[MENU]
Tools
option.
to enter the submenu.
until the
Diagnostics
[OK/MUTE]
to enter the submenu.
option is the first option, press
to enter Sash Sensor diagnostics.
HEIGHT. At the proper sash
when sash 2-3 inches (5-8
to return to the Diagnostic submenu.
option is
AT
157