6. Double check all power has been removed from the biosafety cabinet.
7. Disconnect the two (2) wire harness connections from the motor. The 5-pin
connector, located closest to the blower, is released by depressing the locking
tabs on either side of the connector. The 16-pin data connector is released by
depressing the single locking tab in the center of the connector, on the side of the
connector closest to the blower. See Figure 12-38.
5-pin
Connector
Locking Tabs
Blower
Note: On a Type C1, it may be difficult to reach the two motor wiring connectors
described in Step 7. If this is the case, disconnect the exhaust motor wire harness at
the pass-thru port near the top of the exhaust compartment. Once the exhaust plenum
is free from the biosafety cabinet, the two wire connectors can easily be disconnected
from the exhaust motor. See Figure 12-39.
Figure 12-38
Figure 12-39
126
Motor
16-pin
Connector
Locking Tab