Acceptance Criteria - Labconco Logic+ A2 Technical Manual

Biosafety cabinets, procedure stations
Hide thumbs Also See for Logic+ A2:
Table of Contents

Advertisement

Type
Sash
A2 & B2
Height
Cabinet
Width
inches
8
3-foot
9
10
8
9
4-foot
10
12
8
9
5-foot
10
12
8
9
6-foot
10
12
Sash
Type C1
Height
Cabinet
Width
inches
8
4-foot
10
8
6-foot
10
1.
Tolerance for Average Inflow Velocity = +/- 5 FPM (+/- 0.03 m/s)

Acceptance Criteria

Inflow Volume shall be in the range specified in Table 10-7 column "Inflow Volume
Range".
Inflow
Sash Open Area
1
Average
2
FPM (m/s)
ft
105
2.03
105
2.28
105
2.53
105
2.69
105
3.03
105
3.37
105
4.04
105
3.36
105
3.78
105
4.20
105
5.05
105
4.03
105
4.53
105
5.03
105
6.06
Inflow
Sash Open Area
1
Average
2
FPM (m/s)
ft
105
2.69
105
3.37
105
4.03
105
5.03
Inflow
Volume
2
m
CFM
0.19
213
0.21
240
0.24
266
0.25
283
0.28
318
0.31
354
0.38
424
0.31
353
0.35
397
0.39
441
0.47
530
0.37
423
0.42
476
0.47
529
0.56
636
Inflow
Volume
2
m
CFM
0.25
283
0.31
354
0.37
423
0.47
529
78
Table 10-7
Inflow Volume Range
3
CFM
m
/hr
203-223
345-379
228-251
387-426
253-278
430-472
269-296
457-503
303-333
515-566
337-371
573-630
404-444
686-754
336-370
571-629
378-416
642-707
420-462
714-785
505-555
858-943
403-443
685-753
453-498
770-846
503-553
855-940
606-666
1030-1132
Inflow Volume Range
3
CFM
m
/hr
269-296
457-503
337-371
573-630
403-443
685-753
503-553
855-940