Acceptance Criteria - Labconco Logic+ A2 Technical Manual

Biosafety cabinets, procedure stations
Hide thumbs Also See for Logic+ A2:
Table of Contents

Advertisement

5. Observe the average concentration of aerosol downstream.

Acceptance Criteria

Aerosol penetration shall not exceed 0.01%.
Reinstall the exhaust housing front plate, unsecure temporary Inlet Relief Valve hold
open, reinstall the work surface pieces.
73