Adjusting Blower Speed

During initial or annual certification of the biosafety cabinet, the blower speed may need
to be adjusted. Blower speed must only be adjusted by a trained, qualified certifier. To
adjust the blower speed, following these instructions.
1. From the Main Menu, press [MENU], using
icon. Press [OK/MUTE].
2. Use
[UP]
or
3. From the
Warning
the following screens are restricted.
[DOWN]
to select the
Password
Screen, press
[UP]
or
[MENU]
Password
icon, press [OK/MUTE].
[OK/MUTE]
[OK/MUTE]
to acknowledge the warning that
41
[DOWN]
select the
Tools
WARNING !
RESTRICTED AREA
FOR TECHNICIAN
USE ONLY
Tools