1.
Tolerance for Average Downflow Velocity = +/- 5 FPM (+/- 0.03 m/s)
2. Take the downflow velocity reading at each test point location of the test grid.
3. Average all individual test point velocity readings.
4. If the average is not within the acceptance criteria range, adjust the biosafety
cabinet according to Adjusting the Blower Speed and/or Adjusting the Internal
Damper found in
Acceptance Criteria
Average downflow shall be as specified in Table 10-6 column "Downflow Average" +/-5
FPM (+/-0.03 m/s), and all individual test points within +/-16 FPM (+/-0.081 m/s) from
the average.
Downflow
Test Grid
1
Average
Total Test
FPM (m/s)
Points
55 (.28)
21
55 (.28)
21
55 (.28)
24
55 (.28)
24
60 (.30)
24
55 (.28)
30
55 (.28)
30
60 (.30)
30
55 (.28)
36
55 (.28)
36
60 (.30)
36
Downflow
Test Grid
1
Average
Total Test
FPM (m/s)
Points
65 (.33)
24
65 (.33)
24
55 (.28)
36
55 (.28)
36
Section 9:
Calibration.
Rows x
From Back
Columns
& Sides
3 x 7
6.0 (15)
3 x 7
6.0 (15)
3 x 8
6.0 (15)
3 x 8
6.0 (15)
3 x 8
6.0 (15)
3 x 10
6.0 (15)
3 x 10
6.0 (15)
3 x 10
6.0 (15)
3 x 12
6.0 (15)
3 x 12
6.0 (15)
3 x 12
6.0 (15)
Rows x
From Back
Columns
& Sides
3 x 8
6.0 (15)
3 x 8
6.0 (15)
3 x 12
6.0 (15)
3 x 12
6.0 (15)
75
Table 10-6
Distance
in inches (cm)
Between
Between
Rows
Columns
5.75 (14.6)
4.08 (10.4)
5.55 (14.1)
4.08 (10.4)
5.75 (14.6)
5.21 (13.2)
5.55 (14.1)
5.21 (13.2)
5.44 (13.8)
5.21 (13.2)
5.75 (14.6)
5.39 (13.7)
5.55 (14.1)
5.39 (13.7)
5.44 (13.8)
5.39 (13.7)
5.75 (14.6)
5.50 (14.0)
5.55 (14.1)
5.50 (14.0)
5.44 (13.8)
5.50 (14.0)
Distance
in inches (cm)
Between
Between
Rows
Columns
5.75 (14.6)
5.21 (13.2)
5.55 (14.1)
5.21 (13.2)
5.75 (14.6)
5.50 (14.0)
5.55 (14.1)
5.50 (14.0)
Need help?
Do you have a question about the Logic+ A2 and is the answer not in the manual?
Questions and answers