21: Prefilter Kit

All Type A2 and B2 biosafety cabinets are prepared with the necessary mating
components to accept a Prefilter Kit. To install the Prefilter Kit, follow these instructions.

Kit Catalog Numbers (Type A2 and B2 only)

Installation

Decontaminate all interior work surfaces and work surface components before installing
the prefilter.
1. Remove the work surface of the cabinet by lifting the front edge by one of the
front corner knobs. Slide the work surface forward to disengage the rear tabs.
Support the underneath side with one hand while stabilizing the front edge with
the other. Set the work surface aside. Reference Figure 21-1.
2. Pinch and lift the work surface support straight up to remove it. Make sure it is
surface decontaminated before removing it from the biosafety cabinet.
Reference Figure 21-1.
3. Access the towel catch by pivoting the bottom out towards you. Surface
decontaminate the towel catch before removing it from the biosafety cabinet.
Reference Figure 21-1.
Catalog Number
Cabinet
width (feet)
G4 Paper
3-ft
3406000
4-ft
3406001
5-ft
3406002
6-ft
3406003
187
Table 21-1
Aluminum
3406100
3406101
3406102
3406103

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Logic+ A2 and is the answer not in the manual?

Questions and answers