Setting The Filter Life Gauge; Type A2 (8-Inch Through 10-Inch Sash) And Type B2 - Labconco Logic+ A2 Technical Manual

Biosafety cabinets, procedure stations
Hide thumbs Also See for Logic+ A2:
Table of Contents

Advertisement

Setting the Filter Life Gauge

During initial or annual certification of the biosafety cabinet, and after the HEPA filters
are changed, the Filter Life Gauge may need to be reset. There are three options to
choose from when resetting the Filter Life Gauge. Below are the three options, and an
explanation on which to select.
– Select this option when the HEPA filters are new. This occurs
1.
New Filter
during initial certification of a new biosafety cabinet, or after new HEPA filters are
installed and the biosafety cabinet's Inflow and Downflow have been verified by a
trained and qualified certifier.
2.
Same Filter
Calibration submenu to reach the sensor calibrations, or when the blower
speed(s) have been adjusted slightly (but the HEPA filters have not been
replaced) during an annual certification.
3.
Set Filter Life
Life Gauge to a specific percentage, for example, after the display circuit board is
replaced, and the settings from the previous display board need to be
programmed into the new display board.
To access the Filter Life Gauge screen(s), follow the steps in Adjusting the Blower
Speed previously shown in this section, until on the
these instructions:

Type A2 (8-inch through 10-inch sash) and Type B2

1. On the
Blower Speed
CURRENT RPM = 1217
AVERAGE RPM = 1219
'OK' TO CONTINUE
'OK' displayed when blower speed stabilizes
2. Use
[UP]
or
press [OK/MUTE].
– Select this option when passing through this screen in the
– This option is seldom selected. It may be used to set the Filter
screen, press [OK/MUTE].
58
[DOWN]
to select one of the three Filter Life options displayed,
Blower Speed
[OK/MUTE]
45
screen. Then follow
NEW FILTER
SAME FILTER
SET FILTER LIFE

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Logic+ A2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Logic+ b2Cell logic+ a2Cell logic+ b2Axiom c1Puricare

Table of Contents

Save PDF