Exhaust
Filter Cover
4. Using a 1/2-inch socket, loosen the 4 bolts on top of the cabinet liner (Figure 12-
37), until filter drops 1/2 inch to 3/4 inch (19 mm). Do not remove these 4 bolts,
just loosen them to lower the exhaust filter.
Exhaust Filter Bolts
Exhaust Filter Bolts
5. Slide the exhaust HEPA filter straight out of the biosafety cabinet.
Note: When sliding the exhaust HEPA filter out, the rear edge of the filter frame
may catch and stop the filter from easily sliding out. If this happens, push the
filter back in about 1 inch (25 mm), pull the front edge of the filter down, which
will tilt the back edge up. Maintain this filter position, and slide the filter out.
C1 – ONLY Figure 12-36
Top of cabinet
Figure 12-37
125
Blower Plenum
Cover
Need help?
Do you have a question about the Logic+ A2 and is the answer not in the manual?
Questions and answers