Calibration - Labconco Logic+ A2 Technical Manual

Biosafety cabinets, procedure stations
Hide thumbs Also See for Logic+ A2:
Table of Contents

Advertisement

Calibration

After airflow sensor installation, you MUST calibrate the new sensor. Follow these
steps to calibrate the new sensor.
Note: To properly calibrate the airflow sensor, a qualified certifier must verify the inflow
and downflow values.
1. From the Home Screen, press [MENU].
2. Press
[DOWN]
3. Press
[OK/MUTE]
4. Press the
[DOWN]
highlighted, then press
certifier's submenus.
5. You will see a
are about to alter the BSC's settings; Press
[OK/MUTE]
to input the password.
6. When requested for the password press [LIGHT],
[UV
LIGHT], [TIMER],
until the
Tools
option is highlighted.
to enter the menu screen.
button until the
Password
[OK/MUTE]
Warning Screen
alerting you that you
[OUTLET]
then [OK/MUTE].
option is
to select the
178