Changing the Optional UV Lamp
The UV Lamp is an optional feature and may not be found on all models.
THE UV LAMP IN THIS PRODUCT CONTAINS MERCURY
Manage in accordance with local disposal laws. DO NOT place lamps in trash.
Dispose as a hazardous waste. For information regarding safe handling,
recycling and disposal, consult www.lamprecycle.org
CETTE LAMPE (UV) DANS CE PRODUIT CONTIENT DU MERCUE
Éliminez ou recyclez conformément aux lois applicables. Pour de l'information
concernant des pratiques de manipulation sécuritaires et l'élimination sécuritaire
et le recyclage, veuillez consulter www.lamprecycle.org
For optimum performance, the UV lamp should be changed as indicated by the UV
lamp hourmeter found on the LCD Display in the
The UV lamp and the work area of the cabinet must be thoroughly decontaminated
before removing the lamp.
1. Start the cabinet blower and let it operate for 5 minutes.
2. Raise the sash to its full open position.
3. Thoroughly surface decontaminate the UV lamp and the work area of the
cabinet.
4. Unplug the cabinet from all electrical power.
5. Remove the UV lamp by rotating it 90 degrees and lifting it straight up and out of
its sockets.
6. Install new lamp by reversing the removal procedure.
7. Reset the UV Hourmeter as follows:
In the
Settings
[OK/MUTE]. The hour text will begin to flash, hold
Parameters
submenu enter
UV
Parameters, then select Reset, press
Settings
submenu.
UV
[OK/MUTE]
132
Reset
for 3 seconds.