9. Storage

Avoid long-term storage if possible. When long-term storage is necessary, the pump
should be set up and run for at least eight hours every 12 months.
Use the following procedure to store the pump.
1.
Place protective covers over all ports, the main inlet, inter-stage exhaust, purge
and vent ports.
2.
Place the pump in its packing materials. For fastest pump-down when the pump is
put back into service, seal the pump inside a plastic bag together with a suitable
desiccant.
3.
Store the pump in cool, dry conditions, preferably not exposed to atmospheric air
until required for use. When required, prepare and install the pump as described in
Installation
B8J200880_D
300812671_002_C3
B8J200880_D - Storage
on page 50.