Figure 11 Interface circuits for nEXT turbo pump controllers

4.10 Serial protocol

The serial interface link is set to 9600 Baud, 8 bits, 1 stop, no parity with no
handshaking. The commands are made up from printable ASCII characters. The
maximum message size that can be sent is 80 characters, including start and end
characters.
Note:
All alphabetical characters must be sent in upper case format. The response may contain
lower case characters.