3. Technical data

Note:
The operating, storage conditions and performance of the pumping system may depend
on the type of backing pump used. Refer to the technical data in the appropriate
supplementary publications as listed in

3.1 Operating and storage conditions

Table 1 Operating and storage conditions
Parameter
Ambient operating temperature range
Ambient storage temperature range
Maximum operating humidity
Maximum operating altitude
Sound level
Declared dual number noise emission val-
ues in accordance with ISO4871
Pollution degree (EN61010)
Maximum outlet pressure
▪ XDD1 pump
▪ E2M1.5 pump
Maximum system throughput (N2)
System leak tightness

3.2 Mechanical data

Table 2 Mechanical data
Parameter
Dimensions
Degree of protection (to IEC34-5:1981)
Mass
▪ XDD1 and nEXT ISO100 combination
▪ XDD1 and nEXT CF100 combination
▪ E2M1.5 and nEXT ISO100 combination
▪ E2M1.5 and CF100 combination
D39595880_A
D39595880_A - Technical data
General description
Page 10
on page 8.
Value
Units
12 to +35
°C
-10 to +60
°C
maximum 90
% RH and °C
condensing at 40
2000 maximum
m
56 ± 3
dB(A)
2
-
5
1.1 (1.1 x 10
)
bar absolute (Pa)
5
1.5 (1.5 x 10
)
65
sccm
-6
<1x10
mbar l s
Data
Refer to
Figure: Dimensions
IP20
21 kg
24 kg
26 kg
29 kg
10/2020 - ©Edwards Limited