When BX3M-RLAS-S is combined

Adjusting the analyzer finely
1
If the DIC slider is engaged in the light path, remove it from the light
path. For details, see "4-10 Inserting the DIC slider" (page 62).
2
Engage the 10X objective or 20X objective in the light path and bring
the sample into focus approximately.
3
Insert a stick with a fine point, such as a precision screwdriver, etc. in
the notch a and catch the cover to remove it.
4
Loosen the screw b to unlock.
5
Turn the dial to adjust the analyzer.
Observation with crossed Nichols*: Align the white circle index c
of the dial with the white line d .
Observation without crossed Nichols*: Turn the dial while looking at
the observation image and align it with the position where you can
view the desired observation image.
*
The crossed Nichols means the state that the field of view becomes the
darkest.
During DIC observation, the analyzer rotation dial must be set
TIP
to the crossed Nichols state.
6
Tighten the screw b to lock the dial rotation.
7
Close the cover to return to the original state.