Cause
The cables of the LED lamp housing, motorized nosepiece,
coded nosepiece or coded reflected light illuminator are
not connected.
The brightness of LIM is not stored.
The LED lamp housing is damaged.
The LIM is not set to the Replay mode.
The microscope frame is damaged.
The control box (BX3M-CB) is not connected.
The control box (BX3M-CB) is damaged.
The motorized nosepiece is not connected.
The motorized nosepiece or its cable is damaged.
The hand switch for motorized nosepiece (BX3M-HSRE) is
not connected.
The hand switch for motorized nosepiece (BX3M-HSRE) is
damaged.
The hand switch for motorized nosepiece (BX3M-HSRE) or
the motorized nosepiece is damaged.
The stored brightness (LIM function) is restored to the factory
default setting while the control box (BX3M-CB) is connected.
The power supply was turned ON while pressing the SET
switch of the microscope frame.
Remedy
Set the main switch to
(OFF) and connect the cables of the
LED lamp housing, motorized nosepiece, coded nosepiece and
coded reflected light illuminator to the connector on the back of
the microscope frame.
Operate according to the procedures.
If you cannot adjust the brightness even by pressing the LIM
switch on the front of the microscope frame to set to the Store
mode (The LIM indicator is OFF.), the LED lamp housing is
damaged. Contact Olympus.
Press the LIM switch on the front of the microscope frame to set
to the Replay mode. (The LIM indicator is ON.)
If the LIM indicator does not light by pressing the LIM switch
on the front of the microscope frame, the microscope frame is
damaged. Contact Olympus.
Set the main switch to
(OFF) and connect the control box and
the AC adapter of the control box again.
Then, set the main switch to (ON) and confirm that the beep is
heard once.
If the indicator on the back of the controller does not light even
though the control box and the AC adapter of the control box
are connected again, and also if the beep sound is not heard
when the main switch is set to (ON), contact Olympus.
Set the main switch to
(OFF) and connect the motorized
nosepiece again.
Contact Olympus.
Set the main switch to
(OFF) and connect the hand switch for
motorized nosepiece (BX3M-HSRE).
If the indicator4 on the back of the control box does not turn
ON even though the hand switch for motorized nosepiece is
connected, contact Olympus.
Contact Olympus.
Contact Olympus.
Set the main switch to
(OFF) once and set it to (ON) again.
BX53M
Page
16, 115
27
27
27
27
16, 100
, 117
16, 100
, 117
16, 104
16, 1 15, 100
16
77