Cause
The cable of the coded nosepiece is not connected.
The interface cable is not connected.
The main switch is set to
(OFF) while the
communication between PC and the microscope is in
process.
The device driver is not installed.
The device is not detected properly.
The USB cable is connected to the USB connector of PC
for the first time.
The communication between the control box (BX3M-CB)
and PC is disconnected.
The large supply voltage fluctuation, etc. occurs caused by
thunder, etc. and the system was reset.
The communication between the control box (BX3M-CB) and
the hand switch (BX3M-HS) is disconnected.
The cable of the hand switch (BX3M-HS) is removed while
the power is ON.
The communication between the control box (BX3M-CB)
and the hand switch for motorized nosepiece (BX3M-HSRE)
is disconnected.
The cable of the hand switch for motorized nosepiece (BX3M-
HSRE) is removed while the power is ON.
The communication between the control box (BX3M-CB) and
the hand switch for exposure (U-HSEXP) is disconnected.
The cable of the hand switch for exposure (U-HSEXP) is
removed while the power is ON.
The communication between the control box (BX3M-CB)
and PC is disconnected.
Remedy
Set the main switch to
(OFF) and connect the cable of the
coded nosepiece.
Connect the interface cable.
Close the application software.
Then, set the main switch to (ON) and start the application
software.
Install the device driver from the installation media of the
application software.
Scan the device again with the Windows
Install the device driver in accordance with the installation
wizard.
Remove the interface cable from PC and connect it again. Then,
restart the application software.
If the communication is still not restored, turn OFF the power
supplies of the microscope and PC and turn ON again.
This is not a failure.
If this phenomenon occurs frequently, contact Olympus.
Set the main switch to
(OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
Set the main switch to
(OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
Set the main switch to
(OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
Set the main switch to
(OFF) once and set the main switch to
(ON) again.
If the error occurs frequently, contact Olympus.
BX53M
Page
16, 115
121
16, –
®
device manager.
16, 121
16
16
16
16
83