Olympus BX53M Instructions Manual page 106

System microscope
Table of Contents

Advertisement

Function/Unit
Beep
Motorized nosepiece
Hand switch for
motorized nosepiece
(BX3M-HSRE)
Reserved by maker
100
4
b
c
a
2
1
1
2
3
4
OFF
ON
OFF
ON
OFF
ON
OFF This switch should be always kept OFF.
ON
Attaching the control box
1
Remove the sticker a of the connector on the back of the microscope
frame.
2
Insert the control box (BX3M-CB) by aligning with the attaching section
b on the back of the microscope frame. And push in the control box
until the
mark c is hidden in the attaching section b .
Setting the dip switch
This setting is necessary when combining the motorized nosepiece
or the hand switch for motorized nosepiece (BX3M-HSRE) to the
microscope.
1
Set the dip switch on the side of the control box (BX3M-CB).
The beep sound is heard.
The beep sound is not heard.
5 holes
6 holes
When the right button is pressed, the nosepiece rotates
counterclockwise.
When the right button is pressed, the nosepiece rotates clockwise.
: Factory default setting
Setting