a
1
1
c
B
e
c
2
d
3
This picture shows the procedure to attach the
cable for motorized nosepiece to the microscope
frame. The same procedure applies when attaching
it to the height adapter.
5
Attaching the height adapter
b
1
2
A
6
Attaching the cable for motorized nosepiece
b
a
1
2
e
3
4
4
Use the dedicated Allen wrench (
adapter to attach the height adapter.
Align the mounting screw holes a (4 positions) of the height adapter
with the mounting screw holes b of the microscope frame, and insert
the provided mounting screws c (4 positions).
Depending on the case
A
on the objective you use), while pushing the height adapter to the
microscope frame, tighten the mounting screws c (4 positions) to
secure the height adapter.
When the following objective is not used:
A
Push the height adapter to the back side.
When the following objective is used:
B
Push the height adapter to the front right side.
Applicable objectives:
MPLN50XBD, MPLN100XBD,
MPLFLN50XBD, MPLFLN100XBD, MPLFLN150XBD,
MPLFLN50XBDP, MPLFLN100XBDP
The cable for the motorized nosepiece (BX3M-RMCBL) must
be attached when combining the motorized nosepiece to the
microscope.
Put the cable for the motorized nosepiece through the cable hole a
on the top of the microscope frame and put it out to the back of the
microscope frame.
Align the mounting holes c (2 positions) of the cable for the motorized
nosepiece with the screw holes b (2 positions) on the top of the
microscope frame.
While pushing the metal plate of the cable for the motorized nosepiece
to d positions (2 positions), tighten the provided screw e to secure
the metal plate.
Connect the cable that comes out to the back of the microscope frame
in
1
to the connector of the control box (BX3M-CB). For details, see
Cable connections (page 115).
) provided with the height
or
described below (differs depending
B
BX53M
101

Hide quick links: